This is description of this part:
Set of knobs, shafts and gears for focusing the Olympus CX31 / Olympus CX41 microscopes.
This is the combined system of coarse movement (macro travel) and coaxial fine movement (micro travel). It consists of the following parts:
- left coarse focus knob (manufacturer's name: coarse focus adjustment knob)
- right coarse focus knob (manufacturer's name: coarse focus adjustment knob)
- left fine focus knob (manufacturer's order number: AB059300, manufacturer's name: fine focus adjustment knob (left), reference mark in the manual drawing: A006)
- right fine movement knob with scale (manufacturer's order number: AB467500, manufacturer's name: fine focus adjustment knob (right), reference mark in the manual drawing: A005)
- steel shaft connecting both fine focus knobs (manufacturer's order number: AC732500, manufacturer's name: shaft, reference mark in the manual drawing: A002)
- plastic gear wheel on the fine movement shaft (manufacturer's order number: AA782600, manufacturer's name: gear, reference mark in the manual drawing: A003)
- brass shaft of the coarse movement
- complete reduction gearing (mounted on the casting marked A7829, manufacturer's name: pinion section)
- mount of the fine movement shaft (manufacturer's name: fine shaft mount)
- side cover plates of the knobs (manufacturer's order number: AD071500, manufacturer's name: plate, reference mark in the manual drawing: A001)
- tension ring of the coarse focus knob (manufacturer's name: adjustment ring)
- set of mounting screws
In the list above I do not separately list the assembly that the manufacturer calls "fine focus adjustment knob ass'y" (manufacturer's order number: AQ153900, reference mark in the manual drawing: A004), because it is made up of the parts already named: the shaft, the gear and the right fine focus knob. Olympus also supplied this sub-assembly as a single service part.
Replacing this mechanism can help to solve, among others, the following problems:
- if you feel resistance every now and then while turning the fine or coarse focus knobs. In that case the shaft AC732500 (A002) may be bent
- if the stage sinks down on its own. This symptom may be caused by a damaged or misadjusted tension ring, but note that the same symptom also occurs when the clamping of the vertical guide is loose (https://www.microscopeparts.eu/part/92014)
- if the knobs do not turn at all or turn unevenly. The set contains 5 gear wheels and pinions - damage to any of these parts can produce such a symptom
The order numbers of the parts and the reference marks are given after the manufacturer's original documentation: Olympus Optical Co. Ltd., "Maintenance Manual CX31/41", issued by Marketing Dept., ref. SK9214, November 2001, chapter 5 "List of Maintenance Parts" (http://www.frankshospitalworkshop.com/equipment/documents/microscopes/service_manuals/Olympus%20CX-31,41%20Microscope%20-%20Service%20manual.pdf). The names of the remaining sub-assemblies - after "BIOLOGICAL MICROSCOPES CX31 / CX41 REPAIR MANUAL", Olympus Optical Co. Ltd., Scientific Equipment Division.
The last photos in the gallery show the frame of the Olympus CX31 microscope, with arrows indicating which parts are meant and where they are located in the microscope.
What is this part needed for?