This is description of this part:
Roller guides with brass cages and 16 rollers - a service part of the focus block in the Zeiss ICM 405 / Zeiss IM 35 inverted microscopes.
In the Zeiss IM 35 / ICM 405 inverted microscopes (and of course in the same microscopes sold under the Opton brand) focus is set by raising and lowering the microscope nosepiece together with the objectives. The stage stays still. What I show and offer here is exactly that part of the nosepiece movement mechanism which fails most often: the cages made of thin brass sheet that hold the rollers, and the set of 16 rollers, which often fall out and get lost.
The set contains:
- two side pressure pieces with which the running resistance is adjusted. Loosening of this very element is often responsible for the nosepiece with the objectives sinking down by itself
- two raceways on which the rollers run (name used by the manufacturer: wire guide)
- two brass plates (cages) with rectangular windows, keeping the rollers at even spacing (name used by the manufacturer: casing). These often bend or even tear, which makes the vertical movement impossible or uneven
- a set of 16 steel bearing rollers, eight per side. They often fall out and get lost after a mechanical knock or during servicing of the mechanism
The last photo in the gallery shows the body of the Zeiss IM 35 microscope with the side cover removed, and arrows point to which elements are meant and where they sit in the microscope.
What is this part needed for?
- as a service part, when the brass cages, the pressure pieces or the rollers have to be replaced