More about part: Zeiss ICM 405 microscope mirror with switch from photo camera output to ground glass screen incl. control knob (ID: 99711)
Main Photo
Description
Zeiss ICM 405 microscope mirror with switch from photo camera output to ground glass screen incl. control knobMirror module that diverts the image between the photo camera output and the ground glass screen, together with its microswitch, for the Carl Zeiss ICM 405 inverted microscope.
In the Zeiss ICM 405 inverted microscope the light - that is, the image - which is sent to the photo output can be diverted to the large format ground glass screen. Switching between the photo port and the ground glass is done by sliding the knob on the left-hand side of the microscope body. It moves the internal mirror and at the same time makes and breaks a microswitch which tells the exposure automation where the mirror stands.
What I offer here is exactly this module, with its knob, mirror and microswitch.
The complete module includes, among others, the following parts:
- knob for sliding the mirror
- mirror mounted on slide guides
- microswitch V4T7
- shaped casting that allows the module to be mounted in the microscope body
Anyone who does not use photomicrography on the Zeiss ICM 405 may go for years without knowing that this switch is faulty. It has no effect on the quality of observation through the microscope head - that is, looking with the eye.
Replacing this part can solve, among others, the following problems with the Zeiss ICM 405 inverted microscope:
- when the control knob is missing from the switch. It can break off or work loose and get lost.
- when the mirror has lost its properties, the reflecting surface has oxidised or has been scratched by incorrect cleaning
- when the mirror is mechanically damaged, for example cracked or detached from its holder
- when the microswitch is faulty and no longer sends a signal to the exposure control automation
Note: if you are already sure that it is the microswitch in your microscope that has failed - then you do not have to replace the whole module. It is a V4T7 switch, and equivalents are available on the market today (as of November 2026). One of them is the SAIA-BURGESS V4NCT7.
To replace this module in the microscope the knob has to be unscrewed - otherwise the side cover of the body cannot be taken off. A possible order of work looks like this:
- unscrew the knob of the faulty module
- undo the screws holding the cover and take it off
- remove the faulty module
- unscrew the knob from the new module
- fit the new module
- put the cover back on and screw it down
- screw the knob in
The last photo in the gallery shows the inside of the part of the Zeiss ICM 405 microscope where this mechanism sits. An arrow points to where this mirror module is. In addition, the attached video shows the mechanics of the sliding movement of the mirror and how the microswitch works. This allows the part described here to be identified beyond any doubt.
Manufacturer
Zeiss - (Carl Zeiss West Germany) ➔ All parts
It comes from
This part was removed from the following microscope:
Why do you need this?
This microscope part is typically used for the following purposes:
- as a service part in the form of a complete module replacement, when this assembly or one of its components in the microscope has been damaged
- example faults are listed in the description
Compatibility
This microscope part is compatible with the following microscope models, among others:
- Zeiss ICM 405
- Opton ICM 405